Service Guide 5
Keysight U8001A/U8002A User’s and Service Guide 59
Test consideration
For optimum performance, all procedures should comply with the following
recommendations:
Assure that the calibration ambient temperature is stable and between 20 °C
and 30 °C. Ideally the calibration should be performed at 25 °C ±1 °C.
Assure ambient relative humidity is less than 80%.
Allow one hour warm up period by power on the power supply. During warm
up period, set the power supply to maximum programmable voltage and
current. Then, enable the output.
Keep the test setup connection cables as short as possible.
Resistive Load (R
L
)–10 Ω/90 W (for U8001A)
–6 Ω/150 W (for U8002A)
Measures ripple and noise PV
Current Monitoring
Resistor (Shunt) (R
M
)
–0.01 Ω ± 0.1%
TCR less than 20 ppm/ºC
ISOTEK Co.
Model A-H
Current monitoring
resistor
PV
C
[a] PV = Performance Verification
[b] C = Calibration
Table 5-2 List of equipment (continued)
Equipment Requirement(s) Recommended model Purpose Used