5 Service Guide
62 Keysight U8001A/U8002A User’s and Service Guide
Electronic load
Many of the test procedures require the use of a variable load resistor capable of 
dissipating the required power. Using a variable load resistor requires that 
switches should be used to connect, disconnect, and short the load resistor. An 
electronic load, if available, can be used in place of a variable load resistor and 
switches. The electronic load is considerably easier to use than load resistors. It 
eliminates the need for connecting resistors or rheostats in parallel to handle 
power, it is much more stable than carbon-pile load, and it makes easy work of 
switching between load conditions as is required for the load regulation and load 
response tests. Substitution of the electronic load requires minor changes to the 
test procedures in this chapter.