PLZ-U Application Operation 5-17
5.9 Configuration Settings
Table 5-2 indicates the configuration items.
Table 5-2 Configuration settings
*1. Underlined values are factory defaults.
*2. Up to 3 on the PLZ-30F
■ Procedure
Fig. 5-13 Keys for the configuration settings and display
CONFIG
Number
Item
Selectable Values
*1
Relevant Function
1 Number of units operating in parallel
1
, 2, 3, 4, or 5
*2
Section 5.11, “Parallel Operation”
2 External voltage reference 0, 1, or 2 Section 5.10.2, “CH CONT Connector”
3 Function for turning the load off in the OCP OFF
or ON Section, 4.4.3, “Setting the Trip Points”
4 Function for turning the load off in the OPP OFF
or ON Section, 4.4.3, “Setting the Trip Points”
5 Delayed load-on time setting 0 ms
to 1 s Section 4.2, “Load On/Off”
6 Last load OFF
or ON Section 5.6, “Sequence Function”
7 Polarity of the load-on external control H
or L Section 5.10.2, “CH CONT Connector”
8 Powering up with the load turned on OFF
or ON Section 4.2, “Load On/Off”
9 Communication function selection GPIB
or 232 Chapter 6, “Remote Control”
10 GPIB address 1
to 30 Section 6.4.1, “GPIB Control”
11 RS-232C baud rate 2400, 4800, 9600, or
19
200 bps
Section 6.4.2, “RS-232C Control”