K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -27-
User CRM Calibration_Gas CRM Calibration as shown in Figure 16 below is a user defined
table that allows for the input of flash point values of up to four (4) Certified Reference Material
(CRM) samples using a Gas Ignition source.
1. Input the certified flash point value from the CRM bottle label or test certificate into the first
column titled CRM Flash Value
2. Perform a series of four flash point tests with the analyzer using the CRM sample
3. Input the Flash Point Value as determined by the Analyzer into each Run Column 1 - 4
4. Click the Save Changes button to confirm the instrument calibration
Figure 16: Calibration Screen_Gas CRM Calibration