K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -26-
RTD Correction Table Calibration as shown in Figure 15 below is performed in conjunction with
external certified temperature measuring devices. This table lists 21 fields in which the user can
input offset values for any of temperature values shown. This table is particularly convenient for
those with Certified Digital Contact Thermometers or Certified Liquid in Glass Thermometers and
are interested in offsetting the analyzer at one or more temperature values of their choice. Simply
enter the offset value in the corresponding field and click Save Changes to implement the offset.
Figure 15: Calibration Screen_RTD Correction Table