K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -29-
b. Pressure Calibration - The Flash Point Analyzer is equipped with an Atmospheric Barometric
Pressure Correction Feature. As pictured in Figure 18 below the user can Enter Atmospheric
Pressure in Kilopascals (kPa) into the empty field. The Measured Barometric Pressure is
determined by the analyzers integrated barometer. Clicking Save Calibration Parameters will
enter the Pressure Correction and update the Date Calibrated.
Figure 18: Calibration Screen_Pressure Calibration