K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -30-
c. Igniter Offset Calibration - An Electrical Igniter offset can be entered as a deg Celsius (°C)
value as pictured in Figure 19 below. This method of calibration is useful when a material of
known flash point is tested and the analyzer is displaying a value greater than the known value,
e.g. known value is 53°C, Analyzer displays 55°C. Entering a 2°C value in the field will offset the
analyzer measurement to display 53°C.
Figure 19: Calibration Screen_Igniter Offset Calibration