K71000 Automatic PMCC Flash Point Analyzer
Operation and Instruction Manual
K71000-Manual -35-
12. Factory Settings - The Factory Settings Menu can only be accessed when logged into the Analyzer
under the Factory Login. The Factory settings menu is shown in Figure 25 below. In the Factory
Settings Menu, the user or administrator is capable of adjusting parameters such as the Dip
Mechanism, Lift Mechanism and Electrical Igniter Intensity. Please note that these parameters have
been preset at the Koehler Factory for Optimal performance of the Analyzer.
a. Adjust Dip Mechanism: Pressing the Up and Down Arrows will adjust the level at which the
igniter will dip into the shutter assembly during testing. The Find Dip Home button will
automatically bring the Dip Mechanism to the Home Position (Highest Position), acting as a reset
for making future adjustments. The Check Dip Button will activate the Dip Mechanism for a Test
Dip. The Dip On/Off Switch will activate or deactivate the dipping mechanism.
b. Adjust Lift Mechanism: Pressing the Up and Down Arrows will adjust the level at which the Lift
Mechanism will be in relation to the analyzer base. The Find Lift Home button will bring the lift to
the Home (Lowest Position). The Test Position Button will bring the lift to the Test Position (Low
Position on Test Cup). The Change Position Button will bring the lift to the change position
(Middle). The Clean Position button will bring the lift to the Clean Position (Upper). The Stop Lift
button will automatically stop lift from moving.
c. Set Electrical Igniter Intensity: The Intensity of the Electrical Ignitor can be adjusted as a
percentage both for when the Igniter is intensifying for a Test Dip as well as for when the Igniter
is being used to Re-light the gas flame in the event the flame goes out while using gas ignition.
Enter the desired intensity from (0 - 100) % and click the Save button. The Electrical and Gas
Igniter can also be turned Off or On by pressing the toggle switch shown.