EasyManua.ls Logo

Konica Minolta AeroDR System - Disinfection of the DR Detector

Konica Minolta AeroDR System
140 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
118
6.1 Maintenance and inspection items
Spring connector
If foreign material has adhered to the spring
connectors of the AeroDR Battery Charger2,
and wired cable, remove it with a commercial plastic
brush.
AeroDR Battery Charger2
Spring connector
Wired cable
Spring connector
Wired connection connector
If foreign material has adhered to the wired connec-
tion connectors of the DR Detector and AeroDR UF
Cable, remove it with a commercial plastic brush.
DR Detector
Wired connection connector
AeroDR UF Cable
Wired connection connector
AeroDR Battery Charger
Clean dust on the insert table of the AeroDR Battery
Charger with a soft lint-free cloth dampened with an-
hydrous ethanol or water and wrung well.
DR Detector insert table (side)
DR Detector insert table (front)
AeroDR Battery Charger2
Clean dust on the insert table of the AeroDR Battery
Charger2 with a soft lint-free cloth dampened with
anhydrous ethanol or water and wrung well.
DR Detector insert table
6.1.3 Disinfection of the DR
Detector
If bodily fluid or blood from a patient has contaminated
the surface of the DR Detector, disinfect with a soft lint-
free cloth dampened with a small quantity of the following
disinfectant and wrung well.
• Ethanol for disinfection
• Isopropanol for disinfection
• Commercial chlorine bleach, or 0.5 % hypochlorite
(10-fold dilution of household bleach)
IMPORTANT
•••••••••••••••••••••••••••••••••••••
Bleach and hypochlorite are corrosive, so wash the
bleach o󰀨 well to avoid corrosion.
Be careful not to apply any chemical for disinfection
onto the wired connection connectors and the LEDs.
If you directly apply or spray chemical for disinfec-
tion on the DR Detector, the solution will enter the
instrument through exterior gaps, causing a failure.
Please note that if you use chemical for disinfection
other than the above, it may a󰀨ect the quality, per-
formance, and safety of the DR Detector.
•••••••••••••••••••••••••••••••••••••••••••••••••••••

Table of Contents

Related product manuals