— 22 —
 8. Auto Power Off
  Auto power off function operates when about 30minutes passed after power on this instrument.
  When Auto power off function operates and the instrument powered off, the power-off statue returns to normal by pressing any key.
  And it is possible to release the Auto power off function.
  Turn the function switch from OFF position to the function you want with pressing SELECT key, and power on the instrument.
 9. Battery & Fuse Replacement
DANGER
● Never open the battery cover and the instrument case when making measurement.
● To avoid getting electrical shock, be sure to remove test leads from the instrument when opening the battery cover in order to replace batteries and fuses.
 9-1 Battery replacement
   (1) Remove test leads from the instrument.                    
   (2) Remove the holster from the instrument.
   (3) Loosen one screw on the bottom of the Instrument and open the battery
     Cover then replace batteries.
 9-2 Fuse replacement
   (1) Remove test leads from the instrument.
   (2) Remove the holster from the instrument.
   (3) Loosen two screws on the bottom of the instrument and open the battery
    cover, then replace fuses.
 10. Maintenance
  Use a cloth dipped in water or neutral detergent for cleaning the instrument. Do not use abrasives or solvents.
Fuse : F 600V/10A (Fast action type) φ6.3 x 32mm
F 600V/500mA (Fast action type) φ6.3 x 32mm
Screw
Screw
Battery : Two R6P(AA)1.5V or equivalent