— 15 —
● Standards IEC61010-1, IEC61010-2-033 Measurement CAT 300V, pollution degree 2 / Measurement CAT 600V, pollution degree 2
IEC61010-031
IEC61326-1(EMC)
Environmental standard EN50581 Monitoring and control instruments
● Mode of operation △∑ mode
● Indication LCD maximum value 3999 (ACV/A, DCV/A, Ω, F) / maximum value 5119 (Hz) / units, symbols
● Over range display "OL" symbol is displayed on the LCD.
In case that the value is beyond effective measuring range at the position of Ω function range and manual range.
● Auto-ranging Range shifts to upper range when indicated value is more than 3999.
Range shifts to lower range when indicated value is less than 360.
● Sampling rate approx. 400mS
± 1.6%rdg ± 4dgt
± 2.6%rdg ± 4dgt (50 / 60Hz)
± 3.0%rdg ± 4dgt (〜400Hz)
± 2.0%rdg ± 4dgt (50 / 60Hz)
± 3.0%rdg ± 4dgt (〜400Hz)
± 1.0%rdg ± 4dgt
± 2.0%rdg ± 4dgt
Unspecified
Buzzer beeps below about 70Ω
± 3.5%rdg ± 10dgt
± 3.0%rdg ± 5dgt
± 3.5%rdg ± 5dgt
± 0.1%rdg ± 5dgt
± 2.5%rdg ± 5dgt
010A
(2 Autoranging)
0〜4000uA
(2 Autoranging)
0〜400mA
(2 Autoranging)
0〜10A
(2 Autoranging)
0〜40M Ω
(6 Autoranging)
Test current approx. 0.4mA
400 Ω
〜100uF
(5 Autoranging)
〜10MHz
(8 Autoranging)
Input sensitivity:
〜1MHz / more than 1.5V(RMS.)
over 1MHz /
more than 2V(RMS.)
0.1〜99.9%
(Pulse width / Pulse period)
4A
10
400 uA
4000 uA
40 mA
400 mA
4A
10
400 Ω
4kΩ
40 kΩ
400 kΩ
4MΩ
40 MΩ
40 nF
400 nF
4uF
40 uF
100 uF
5.12 Hz
51.2 Hz
512 Hz
5.12 kHz
51.2 kHz
512 kHz
5.12 MHz
10 MHz
DUTY
DCA
ACA
Capacity
Frequency
Resistance
Diode check
Continuity check
Allowable measurement time : 10A(15seconds max)
Allowable measurement time : 10A(15seconds max)
Ω