Instruction manual for CoolSafe
06012017QA01
Item No. 700749
CopyRight © 2017
5.3 Preparations
Check that the CoolSafe ice condenser is free from ice and water, and that all valves i.e. the drain
valve is closed. Check that there is no water in the drain valve.
Note
Do not use any tools to close valves, as they will easily be damaged.
Check that the gas ballast valve on the vacuum pump is open at all times.
Check the oil level in the pump and the oil mist filter. The filter should have only a maximum
10 mm oil level (preferably it should be empty).
Switch on the CoolSafe and the vacuum pump.
The lowest temperature should be achieved within 60 minutes.
The vacuum pump should now be allowed to run for at least 30 minutes to achieve the
lowest vacuum performance. The electrical vacuum valve remains closed until you enter
the freeze-drying modes (opened via the green switch button on the front of the Basic or
Pro cabinets).
5.4 Freeze drying using manifolds
1. The manifold is connected to the conus in the lid. Make sure this connection is tight. Check
that all rubber valves are closed (A) or in vacuum-release position (C), and that the flange
and lid are tight.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LaboGene CoolSafe Basic 110-4 and is the answer not in the manual?

LaboGene CoolSafe Basic 110-4 Specifications

General IconGeneral
BrandLaboGene
ModelCoolSafe Basic 110-4
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals