leak from the drain-this can be done placing a finger over the drain spigot, you will feel if
there is a leak.
4.
Check the seal on the acrylic plate of the chamber that interfaces with the
top stainless steel
part of the CoolSafe. This can be done by squirting or brushing some frothy soap solution
all around, if there is a leak this will be sucked in towards the condenser.
5. If leak(s) are found on the seals, clean both the surface of the s/steel and the seal, using a
clean tissue slightly wetted with alcohol.
6. Check that the lid of the chamber is located correctly, again check that the seal is OK as
previous.