Integral Process Thermostats and High-Temperature Thermostats
Operation manual
IN 130 T, IN 230 T, IN 230 TW, IN 530 T, IN 530 TW, IN 1030 T, IN 1330 TW, IN 1830 TW, IN 150 XT,
IN 250 XTW, IN 280 XT, IN 280 XTW, IN 550 XT, IN 550 XTW, IN 590 XTW, IN 750 XT, IN 950 XTW, IN 1590 XTW,
IN 1850 XTW, IN 2560 XTW, IN 2050 PW, IN 2560 PW, IN 4 XTW, IN 8 XTW
V6R9
Read this manual prior to performing any task!

Table of Contents

Questions and Answers

  • W
    William WrightSep 10, 2025
    Why is degassing functioning poorly on my Lauda Thermostat?
    • M
      Miss Sandra SmithSep 10, 2025
      Degassing on your Lauda Thermostat might function poorly if pressure control is active, so switch it off. A pump level that is too high can also cause this, so select a lower pump level. Reduce the heating output, as it may be too high. If the cooling unit is active, switch it off. If the heat transfer liquid is heavily contaminated, change it. Ensure the filler nozzle is not closed by removing the tank lid and placing it loosely on the filler nozzle. If applicable, connect the inert gas blanket, with compressed air for pure degassing if necessary.
  • D
    Dale BowersSep 12, 2025
    What does 'Pump blocked' mean on Lauda Thermostat display?
    • J
      Joshua CastilloSep 12, 2025
      The 'Pump blocked' message on your Lauda Thermostat display indicates a pump motor monitoring issue, such as overload or blockage. The viscosity of the heat transfer liquid may be too high; consider changing to another heat transfer liquid or increasing the set temperature.
  • J
    John BarnesSep 14, 2025
    What does 'Evaporator frozen' mean on Lauda Thermostat display?
    • J
      Jennifer WalkerSep 14, 2025
      The 'Evaporator frozen' message on your Lauda Thermostat display indicates that the evaporator is frozen. Increase the pump level. Open the manual bypass in the constant temperature equipment. Increase the flow rate through the application.
  • M
    Melanie GibsonSep 16, 2025
    What does 'Low level (pump)' mean on Lauda Thermostat display?
    • K
      Kenneth BrownSep 16, 2025
      The 'Low level (pump)' message on your Lauda Thermostat display indicates that the pump motor is idling. Check whether the float in the expansion vessel is blocked by a foreign object.

Summary

Safety

Safety information

Comprehensive safety guidelines and warnings for device operation.

Before starting up

LAUDA heat transfer liquids

Information on recommended fluids for optimal and safe device operation.

Commissioning

Switching on the device for the first time

Initial power-up sequence and essential configuration steps.

Operation

General warning

Crucial safety advisories for device operation.

Operation of a high-temperature thermostat

Safety instructions for high-temperature thermostats with water counter-cooling

Specific safety advice for high-temperature models with water cooling.

Operating a device with pressure overlay

Warnings, safety instructions and further information

Key safety and operational advice for pressure overlay systems.

Maintenance

Faults

Troubleshooting

Guide to diagnosing and resolving common operational problems and errors.

Lauda IN 280 XT Specifications

General IconGeneral
BrandLauda
ModelIN 280 XT
CategoryThermostat
LanguageEnglish