EasyManuals Logo

Lauda RE 112 User Manual

Default Icon
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
Pre
p
arations
8
Emptying
Switch off the thermostat, pull out the mains plug!
Drain the bath liquid through the drain cock
!
using tubing.
The units are designed for operation with non-flammable and flammable liquids to
EN 61010-2-010! Flammable liquids can be operated up to no more than 25°C
below the firepoint (
Section 5.3.).
Observe the appropriate regulation when disposing used thermostating liquid.
When connecting an external consumer take care of the bath liquid level for it
must not decrease too much
!
fill in bath liquid if necessary.
Do not drain the thermostating liquid when it is hot or very cold (below 0°C)!

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RE 112 and is the answer not in the manual?

Lauda RE 112 Specifications

General IconGeneral
BrandLauda
ModelRE 112
CategoryThermostat
LanguageEnglish

Related product manuals