Startin
g
u
p
20
6.5.4 Other error messages
Other error messages shown on display.
EEE
changes with
error code, e.g. 0
!
various temperature probe faults.
!
pump fault, proceed as in Section 6.5.3.
If the fault report is repeated
!
pull out the mains plug and try whether the motor
can be rotated by the fan blade inserting a screwdriver into the ventilation opening
at the back of the unit.
Error code 0 ...255
!
microprocessor error.
Error code 1000...1255
!
slave processor error.
Indication can be used for remote diagnosis.
After rectifying the fault, reset with the key.