Preparations
13.04.99
YACE0059
7
The unit can be operated safely up to an ambient temperature of 40°C
Depending on the loading of the refrigeration system, a temporary shut-off can
occur, especially in case of an ambient temperature of over 35°C.
Additionally a higher ambient temperature results in less refrigerating capacity.
When starting up the refrigeration system after a longer time, it can take up to 30
min, depending on the ambient temperature and the unit type, until the nominal
refrigerating capacity is reached.
5.2 Filling and emptying
Filling
Close the drain cock.
Fill baths up to a maximum level of 20 mm below the bath
bridge.
Optimum operation at 20-40 mm below the bath bridge.
Operation is possible down to 60 mm below the bath bridge.
The low-level cut-out operates approx. 90 mm below the
bath bridge!
When using thermal oils it is necessary to allow for an expansion of approx.
8 %/100 °C.
When operating with an external consumer the total expansion takes place in the
bath!