EasyManuals Logo

Lauda RE 112 User Manual

Default Icon
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Startin
g
u
p
18
"
""
"

When the bath temperature arises above the
overtemperature switch-off point.
1. Double signal tone sounds
2. The display shows the indication for overtemperature
ttt
the
fault triangle is flashing
!
heating is switched off on both poles,
!
pump and refrigeration system are switched off by the
electronics.
Rectify the cause of the fault.
Wait until the bath temperature has cooled down below the
switch-off point or set the switch-off point at a value higher
than the bath temperature. then the display shows
ttt
Reset with the key
Before the unit is running unattended for longer periods
overtemperature protection
should be tested.
Therefore
Turn the potentiometer slowly anticlockwise.
!
The unit
must switch off at the bath temperature.
Step 1 - 2 (see above) must follow.
Set the overtemperature switch-off point again above the
bath temperature and wait until the indication
ttt
appears on
the display, then
reset with the key.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lauda RE 112 and is the answer not in the manual?

Lauda RE 112 Specifications

General IconGeneral
BrandLauda
ModelRE 112
CategoryThermostat
LanguageEnglish

Related product manuals