EasyManua.ls Logo

Launch KES-200 - Page 32

Default Icon
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LAUNCH KES-200 Starting System Analysis
E-8
Fig.4-17
Parallel wave: It is to display waveforms of several
cylinders in parallel, as shown in Fig.4-18.
Fig.4-18
Bar graph: It is to display the maximum voltage of
each cylinder in the form of the bar graph and value.
The bar graph of secondary ignition waveform is as
shown in Fig.4-19.
Fig.4-19
Parade wave: It is to display the waveforms of several
cylinders in parade mode, as shown in Fig.4-20.
Fig.4-20
8. [7. Positive/ Negative]
It is to select the polarity of discharge, and only
applicable to engines with independent ignition
system or dual cylinder ignition system.
9. [0.GRID] -- To display or hide grid on the screen
10. [Current Page] – To view the currently saved
waveform in the extended function menu
KES-200 can automatically save up to 50 pages of
waveforms and stop when pressing [0.Hold]. The user
can press [3.Current Page] and then use [] []
keys to view the current waveform.
11. [7.SAVE] --To save waveforms
The function is the same as that on the oscilloscope
mode. Up to 50 pages of waveform data can be
automatically saved by KES-200. But the waveform
data will be cleared when the user exits the ignition
waveform function. To keep the waveforms for a long
time, the user must save the waveforms in the internal
memory of KES-200. The user can download the
saved waveform to PC for further analysis.
To save the data, select the memory address, and
then press [ENTER] key. The address with saved
data will be marked with the sign “ * ”.
12. [8. RCLL] --To recall saved waveform data
KES-200 has 8 addresses for storing the waveforms.
When [8.RCLL] is pressed, the eight memory
addresses will be listed. The address with saved data
is marked with the sign “ * ”.

Table of Contents

Related product manuals