EasyManuals Logo

LAUREL LAUREATE 2 Series User Manual

LAUREL LAUREATE 2 Series
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #23 background imageLoading...
Page #23 background image
- 23 -
Option board dependent menu items
ALSEt. ALS34 dEU1H dEU2H dEU1b dEU2b dEU3H DEU4H DEU3b DEU4b
Menu items related to alarm setup These will only appear if a relay board is detected. If so,
see Section 16.
AnSEt. An Lo. An Hi..
Menu items related to analog output setup. These will only appear if an analog output
board is detected. If so, see Section 17.
SEr 1. SEr 2. SEr 3. SEr 4. _Addr
Menu items related to communications. These will only appear if an RS232, RS485, USB
or Ethernet board is detected. If so, see Section 18.
Menu lockout items
Loc 1. Loc 2. Loc 3.
Menu items used to enable or lock out (hide) other menu items. Loc menu items may in
turn be locked out by a hardware jumper on the power supply board. See Section 9.
How to Maximize Accuracy of Load Cell Measurements
1. Avoid plated connectors or terminal blocks, since dissimilar metals can set up thermo-
electric voltages in the millivolt range. If possible, use copper wiring throughout. If you have
to use a junction box, have copper wire touch copper wire.
2. With long cable runs, use six-wire hookup, where the extra two wires are used to sense the
excitation voltage at the load cell. This allows ratiometric meter operation to correct for
voltage drops in the excitation wires.
3. Use twisted, shielded wire from the meter to the load cell to minimize pickup of electrical
noise. Connect the shield to signal input ground at the meter. Do not connect the other end
of the shield.
4. Do not run low-level signals in the same cable tray as electrical power to minimize pickup
of electrical noise carried by power lines.
5. If pickup electrical noise is unavoidable, use the digital filtering capability of our meter to
apply a time constant that averages out random noise over time.
6. Use 5V excitation, not 10V excitation, with miniature or button-type load cells. Even though
small 350-ohm bridge load cells may be rated for 10V, the higher voltage will produce four
times as much self-heating and will cause thermal errors.
7. Operate load cells within their linear range. Do not exceed specified maximum operating
limits so as not to damage the load cell.
8. Operate load cell along their principal axis. Do not apply lateral or twisting forces, as these
will cause deformation and produce non-intended electrical signals.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the LAUREL LAUREATE 2 Series and is the answer not in the manual?

LAUREL LAUREATE 2 Series Specifications

General IconGeneral
BrandLAUREL
ModelLAUREATE 2 Series
CategoryMeasuring Instruments
LanguageEnglish