Instructions for Use
Leica EG1150 H
Paraffin Embedding Station
Leica EG1150 H V 2.6, English 07/2016
Order No.: 14 0388 81101 RevI
Always keep this manual with the instrument.
Read carefully before working with the instrument.

Other manuals for Leica EG1150 H

Questions and Answers

  • Q
    qfranklinAug 9, 2025
    What to do if paraffin does not melt in Leica EG1150 H Laboratory Equipment?
    • R
      Ray AndersonAug 10, 2025
      If the paraffin in your Leica Laboratory Equipment isn't melting or is melting very slowly, consider these possible causes and solutions: * Increase the temperature for the paraffin reservoir. * Activate the on-demand heater. * Check the temperature settings and start time, and repeat if necessary. * Check the fuses and replace them if required. * Keep the dispenser open at maximum flow for a period of time. If these steps don't resolve the issue, it may indicate a technical fault, and you should wait until paraffin is entirely molten then check again.
  • C
    Christina GibsonAug 16, 2025
    What to do if Leica EG1150 H shows an error message and loses program/time?
    • L
      Lisa FloresAug 16, 2025
      If your Leica Laboratory Equipment displays an error message and loses the program and time, it likely means the battery is empty. This can also happen due to a power failure, if the device was switched off with the standby switch, or if the power plug was disconnected. To address this: * Acknowledge the fault by pressing any key. * Reprogram the device to resume work. Important: Avoid disconnecting the instrument from the power supply again. It is recommended to ask technical service to replace the battery.
  • J
    josephmartinAug 25, 2025
    Why does Leica EG1150 H display a flashing colon and not work?
    • E
      ejacksonAug 25, 2025
      If your Leica Laboratory Equipment displays a flashing colon and isn't working, it's likely in the heating phase (standby mode). This is normal; the instrument will switch to operating mode at the programmed start time. Ensure the standby switch is ON, check the instrument's connection, and try holding down the ON/OFF button for several seconds (at least 2 seconds). It may also indicate that the standby switch is not turned on or the cut-out function of the standby switch has been activated or the ON/OFF button has not been pressed long enough.
  • T
    tcollinsSep 1, 2025
    What to do if illumination doesn't work on Leica EG1150 H?
    • M
      Melanie RichSep 1, 2025
      If the illumination on your Leica Laboratory Equipment isn't working, try the following: * Press the On/Off button on the head of the lamp. * If that doesn't work, the bulb may be defective and need replacing by technical service. * Also, check the fuse and change it if necessary.

Summary

Safety

General Safety Notes

Provides essential safety precautions and warnings for using the instrument.

Instrument Warnings

Highlights specific warnings related to operation, transport, and installation hazards.

Instrument Components and Specifications

Instrument Setup

Unpacking and Installation Procedures

Step-by-step guide for safely unpacking and initially installing the instrument.

Necessary Assembly Work

Instructions for assembling accessories and preparing the instrument for use.

Electrical Connection Guide

Instructions for safely connecting the instrument to the power supply.

Operation

Paraffin Reservoir and Dispenser Operation

Details the paraffin reservoir's capacity and temperature control, and the dispenser's flow adjustment.

Control Panel Functions

Overview of the control panel buttons, display, and LED indicators.

Maintenance and Cleaning

Troubleshooting

Addressing Common Faults and Errors

Lists common problems with display, operation, and illumination, with solutions.

Resolving Reservoir, Dispenser, and Fuse Issues

Solutions for paraffin melting, dispenser flow, error messages, and changing fuses.

Leica EG1150 H Specifications

General IconGeneral
TypeParaffin Embedding Center
WeightApprox. 37 kg
Voltage230 V
Paraffin Tank Capacity3.8 L
Cooling Plate Temperature-5°C to +5°C