IFU-0002 Lenire User Manual Version 1.0 Page 49
Chapter 25 - Guidance and Declaration for Electromagnetic Compatibility
The Lenire system has been tested for electromagnetic performance to
the following standards:
» IEC61000-4-3 (radiated electromagnetic fields)
80 MHz to 2.7 GHz at 10V/m
2.7 GHz to 6 GHz at 3V/m
Spot frequencies per EN 60601-1-2:2015 Table 9 at up to 28 V/m
» IEC 61000-4-2 (Electrostatic Discharge)
+/- 8kV Contact Discharge
+/- 15kV Air Discharge
» IEC 61000-4-8 (50Hz/60Hz magnetic fields) at 30 A/m
Disturbances higher than these limits are unlikely to be encountered in the
home healthcare or professional healthcare environments.