F. EMC supplement
F.1 General
The Lenstar LS 900 complies with the standard EN 60601‑1‑2. The instrument is
constructed such that the generation and emission of electromagnetic interference
is limited so as not to disrupt the intended use of other devices, while the instrument
itself possesses an appropriate immunity to electromagnetic interference.
WARNING!
• Electrical medical devices and systems are subject to special EMC mea‑
sures and must be installed in accordance with the EMC instructions con‑
tained in this accompanying document.
• Portable and mobile HF communication systems may interfere with elec‑
trical medical devices.
• The use of cables or equipment other than those listed may lead to a
higher emission or to reduced interference immunity of the Lenstar LS
900 biometer system.
F.2 Emitted interference (standard table 1)
This product is intended for use in the electromagnetic environment specifi ed below. The customer or the user of this product should assure that it is used in such an envi-
ronment.
Emission test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 This product uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions CISPR 11 Class B This product is suitable for use in all establishments, including domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings used for domestic purposes.
Emission of harmonics
according to EN 61000-3-2
Class A