Questions and Answers

  • A
    Aaron GrayAug 14, 2025
    What to do if my LEYBOLD Water Pump does not start?
    • D
      Darren HughesAug 14, 2025
      If your LEYBOLD Water Pump isn't starting, several issues could be the cause. First, check the wiring and repair it if faulty. For three-phase motors, ensure the motor protection switch is properly set up. Also, verify that the power supply voltage matches the motor's rating. Other potential causes include a faulty motor (exchange it), oil temperature being less than 12 °C (warm up the pump and oil or use a different oil), oil being too thick (change the oil), a clogged exhaust filter or line (exchange the filter or clean the line), or a seized pump (repair the pump).
  • D
    Douglas Austin PhDAug 17, 2025
    Why is my LEYBOLD D5E getting hotter than usual?
    • C
      Catherine DavenportAug 17, 2025
      If your LEYBOLD Water Pump is overheating, several factors could be responsible. Check if the supply of cooling air is obstructed and ensure the pump is properly installed. High ambient temperature or process gas that is too hot can also cause overheating, so ensure the pump is properly installed and change the process if needed. Insufficient oil levels or unsuitable oil types could be the cause, so fill in or change the oil. Other potential issues include obstructed oil circulation (clean or repair the oil channels), a clogged exhaust filter or line (change the filter or clean the line), a faulty exhaust valve (repair the valve), or deviating mains supply voltage (check the power supply rating of the motor and the mains supply).
  • J
    Jill NelsonAug 21, 2025
    What to do if my LEYBOLD D5E Water Pump is extremely loud?
    • D
      Diane ElliottAug 21, 2025
      If your LEYBOLD Water Pump is making excessive noise, start by checking the oil level and top it up if it's too low. High intake pressure can also cause noise, so reduce it if necessary. Other potential causes include a clogged internal demister (clean or replace it), worn coupling bushings (replace them), damaged vanes or bearings (repair the pump), or a faulty motor bearing (exchange the motor bearing).
  • M
    Michael WatersAug 25, 2025
    Why LEYBOLD D5E does not attain its ultimate pressure?
    • M
      martinezcarlaAug 25, 2025
      If your LEYBOLD Water Pump isn't reaching its ultimate pressure, consider these possible causes. Ensure you're using the right method of measurement or measuring instrument. Check for external leaks by measuring the pressure directly at the intake port of the pump. Other potential issues include a dirty internal demister or faulty exhaust valve (repair the pump or valve), unsuitable oil (change and possibly degas it), a dirty or leaky intake line (clean or exchange it), or the pump being too small (check process data and possibly change the pump).
  • B
    Benjamin HarrisAug 27, 2025
    What to do if pumping speed of LEYBOLD D5E Water Pump is too low?
    • J
      Jose ThompsonAug 28, 2025
      If the pumping speed of your LEYBOLD Water Pump is too low, check if the dirt trap in the intake line is clogged and clean it if necessary. Additionally, a clogged exhaust filter can reduce pumping speed, so exchange the filter element. Finally, ensure that the connection lines are sufficiently wide and short.
  • D
    Diana OliverAug 30, 2025
    What to do if there is oil in the intake line of my LEYBOLD D5E?
    • V
      Vincent GarciaAug 31, 2025
      If you find oil in the intake line or vacuum chamber of your LEYBOLD Water Pump, it could be coming from the system itself, so check the system. Another possibility is a blocked tandem valve, in that case clean or repair it.
  • A
    Andrea CaseySep 3, 2025
    What to do if after switching off LEYBOLD D5E the pressure increases too fast?
    • M
      Michelle CoxSep 3, 2025
      If the pressure in the system increases too fast after switching off your LEYBOLD Water Pump, it could indicate a system leak, so check the system. Other potential causes include a faulty tandem valve, inner shaft seal, or exhaust/bypass valve. In such cases, repair the valve(s) or exchange the shaft seal.
  • M
    Mitchell RussellSep 6, 2025
    Why is the oil in my LEYBOLD Water Pump turbid?
    • T
      Tracy SheltonSep 6, 2025
      If the oil in your LEYBOLD Water Pump is turbid, it's likely due to condensation. You should degas the oil or change the oil and clean the pump.
  • E
    Ernest HesterSep 9, 2025
    Why is there an oil leak at the side of the motor of my LEYBOLD D5E?
    • A
      Andrew JonesSep 10, 2025
      If you notice an oil leak at the side of the motor of your LEYBOLD Water Pump, it's likely due to a damaged or worn-out outer shaft seal. You should exchange the shaft seal.

Summary

IMPORTANT SAFETY CONSIDERATIONS

1 Description

1.1 Principle of Operation

Explains the working mechanism of the dual-stage rotary vane vacuum pump.

1.4 Technical Data

Presents detailed technical specifications for TRIVAC E models, including performance and dimensions.

2 Operation

2.1 Installation

Guides on how to properly install the TRIVAC E vacuum pump.

2.3 Electrical Connection

Details the electrical connection procedures for AC and three-phase motors.

2.5 Operation Modes

Guidance on pumping different gas types and optimal operating temperatures.

2.6 Shut Down and Shelving

Procedures for safely shutting down and storing the TRIVAC E vacuum pump.

3 Checking and Maintenance

3.1 Checking the Oil Level

Instructions for checking the oil level and its condition visually, chemically, and mechanically.

3.2 Oil Change

Step-by-step instructions for changing the pump oil, including disposal.

3.5 Electric Motor Disassembly and Assembly

Procedures for disassembling and reassembling the electric motor.

4 Maintenance Plan

5 Maintenance and Repair Kits

5.1 Ordering Information for Maintenance and Repair Sets

Lists part numbers for ordering maintenance kits and repair sets for TRIVAC E pumps.

6 Troubleshooting

EC Declaration of Conformity

Declaration of Contamination