Status Registers (Group Address = 0x0060)
Address Monitored
Parameter
Type/Unit Description
10= EEPROM_WRITE_ VERIFY_WARNING
Resettable Warning
9= EEPROM_TIMEOUT_ WARNING
Resettable Warning
8= EEPROM_INVALID_ NUMBER_OF_BYTES
Resettable Warning
7= FLASH_MEMORY_ CRC_BOOT_ERROR
Non-resettable fault detected by the boot loader -
need to repeat the FLASH download process.
6= FLASH_WRITE_ PROTECT_FAULT
Non-resettable fault detected by the boot loader -
need to repeat the FLASH download process.
5= FLASH_CAL_DATA_CRC_ERROR
Non-resettable fault – if fault is not cleared by a
power cycle the Processor board must be replaced.
4= EXTERNAL_RESET
Power on self test warning – generates a fault log
and then automatically clears the fault and restarts.
3= SINK_TEMPERATURE_ SENSOR_FAULT
Non-resettable fault. The software monitoring of the
sink temperature sensor indicates that this sensor is
not working correctly.
2= EEPROM_COMBINED_ GAIN_OF_ZERO
Non-resettable fault
1= EEPROM_INVALID_ BLOCK_CRC
Non-resettable fault
0(LSB)= EEPROM_INVALID_ MAP_ID
Non-resettable fault
0x006A Diagnostic
Register 02
15(MSB)= Reserved
14 = Reserved
13 = Reserved
12= Reserved
11= Reserved
10= Reserved
9= Reserved
8= Reserved
7= PLL Loss of Lock. Phase Locked Loop Loss of
Lock
6= VOLTAGE_DOUBLER _FAULT
Set to indicate that the voltage doubler is not in the
state commanded (detected from measurement of
mains input voltage and Vdc). Note that the raw
state of the voltage doubler relay output is set in the
300902516_002 - 10/2020 - © Leybold
42