ECODRY plus
Multi-stage Roots vacuum pump
Operating Instructions 300758785_002_C1

Questions and Answers

  • B
    brendajonesAug 4, 2025
    What to do if LEYBOLD ECODRY Plus does not reach nominal speed 210 Hz?
    • B
      Barbara DiazAug 4, 2025
      If the LEYBOLD Water Pump does not reach the nominal speed of 210 Hz, it could be due to: * Inlet pressure exceeding 50 mbar. Close the pump inlet and check the rotational speed on the display. * Blocked outlet or unacceptably high back pressure in the outlet. Ensure the pump outlet is clear and that the outlet pressure does not exceed 1200 mbar.
  • D
    Deborah GriffinAug 9, 2025
    Why LEYBOLD Water Pump does not start?
    • I
      iruizAug 9, 2025
      The LEYBOLD Water Pump might not start due to several reasons: 1. No mains voltage present. Check the mains power supply and ensure the mains switch on the pump is set to '1'. 2. Defective or incorrectly inserted mains cable. Verify the pump’s power supply. 3. Incorrect connection of the start-stop switch. If operating using the start-stop switch, ensure the supplied plug is connected to the X1 control interface with pins 8 and 9 bridged. 4. Start-stop button is not set to 'start'. If operating using the X1 control interface, set the start-stop switch to 'start'. 5. DC Under voltage or Controller Under voltage. Compare the voltage range on the pump’s nameplate with the available mains voltage and connect the pump to a power supply at the correct voltage.
  • N
    noahkelleyAug 15, 2025
    What to do if LEYBOLD ECODRY Plus Water Pump end pressure will not be reached?
    • T
      Tim KruegerAug 15, 2025
      If the end pressure is not reached in your LEYBOLD Water Pump, consider the following: * The gas ballast valve might be open. Close the gas ballast valve, if possible. * There could be a leakage in the system. Check for any leaks. * The gas ballast valve could be defective. Check for leaks. * The outlet might be blocked, or there's unacceptably high back pressure in the outlet. Ensure the pump outlet is clear and that the outlet pressure does not exceed 1200 mbar. * The pump might not be running at the rated rotational speed of 210 Hz. Set the target speed to 210 Hz using the X104 RS-485 interface.
  • D
    Danielle GonzalezAug 22, 2025
    Why LEYBOLD ECODRY Plus switches off?
    • G
      goodwinelizabethAug 22, 2025
      The LEYBOLD Water Pump may switch off due to: * Heatsink Overheat. Set the pump up at a suitable location with a lower ambient temperature. * Cooling air supply or discharge impeded. Ensure sufficient cooling air supply or discharge. * Gas temperature too high. Check the system.
  • A
    Audrey VasquezAug 27, 2025
    What to do if LEYBOLD Water Pump does not start or switches off?
    • R
      Robert ReeseAug 27, 2025
      If your LEYBOLD Water Pump does not start or switches off, it could be due to: * Voltage too low for controlling the frequency converter. Disconnect the pump from the mains for at least 10 seconds and restart. * Charging circuit for DC bus is defective. Disconnect the pump from the mains for at least 10 seconds and restart. * DC bus voltage was too high, the rundown time is too short, or the input voltage is too high. Compare the voltage range on the pump’s nameplate with the available mains voltage and connect the pump to a power supply at the correct voltage.

Summary

Important Safety Information

Mechanical Hazards

Details hazards related to mechanical aspects of the pump and safe handling.

Electrical Hazards

Covers electrical safety precautions and requirements for the pump.

Thermal Hazards

Warns about burn risks from hot pump components.

Hazards Caused by Materials and Substances

Discusses risks from materials used and substances pumped.

Noise Hazard

Description

Supplied Equipment

Lists items included with the pump upon delivery.

Technical Data

Provides detailed technical specifications and parameters of the pump.

Electrical Data

Ordering Data

Accessories

Transport and Storage

Transport

Instructions for safely lifting and moving the pump during transportation.

Storage

Guidelines for proper storage conditions and duration to prevent damage.

Installation

Placement

Instructions on positioning the pump on a level surface for optimal cooling.

Conforming Use

Defines the intended gases and conditions for safe and effective pump operation.

Non-conforming Use

Connecting the Pump

Exhaust Line

Electrical Connection

Operation

Manual Operation

How to operate the pump using the physical controls.

Interfaces

Overview of available interfaces for pump control and monitoring.

X1 Control Interface

X104 Serial RS-485 Interface

Frequency Converter Display

Gas Ballast

Swapping out the Gas Ballast Switch

Switching off; Decommissioning

Service

Servicing

Recommends precautionary service intervals and maintenance.

Leybold Service

Procedures for sending equipment to Leybold Service and contamination declarations.

Troubleshooting

Faults and Alarms Overview

Explains how alarms and faults are indicated and how to reset them.

Disposal; Certificates

Waste Disposal

Guidelines for decontaminating and disposing of the pump and its parts.

Certificates

Lists the certifications and testing standards the pump has met.

EU Declaration of Conformity

Declaration of Contamination

LEYBOLD ECODRY Plus Specifications

General IconGeneral
BrandLEYBOLD
ModelECODRY Plus
CategoryWater Pump
LanguageEnglish

Related product manuals