3 Installation
3.1 Placement
Place the pump on a straight and level surface (max. tilt of 2°).
To ensure sufficient cooling of the pump, the fan openings of the pump must
not be obstructed, see fig. 3.2.
3.2 Conforming Use
The ECODRY plus has been designed to produce a vacuum. It is capable of
pumping clean, dry gases, which are free of particles, like air, nitrogen, oxy-
gen ≤ 21%, argon with inlet pressure < 100 mbar and helium. Moreover, it is
capable of evacuating vessels of vacuum systems down to the medium vac-
uum range.
The pump is suited for deployment as a backing pump for turbomolecular
pumps.
Organic solvents may only be pumped in traces/at low concentrations (con-
sult Leybold first) outside the flammability range. The pump is suited for
pumping water vapour within the limits given in the technical data.