After removing a glucose test strip from its vial, immediately close
the vial cap tightly.
Touch the glucose and ketone test strips with clean and dry hands.
Use each test strip immediately after removing it from the vial.
Write the opening date on the vial label when you first opened it.
Discard remaining blood glucose or β-ketone test strips after 6
months.
Do not use test strips beyond their expiry date. This may cause
inaccurate results.
Do not bend, cut, or alter a test strip in any way.
Keep the strip vial away from children since the cap and the test
strip may be a choking hazard. If swallowed, promptly see a doctor
for help.
For further information, please refer to the test strip package insert.
Important Control Solution Information
Use only Our control solution with your meter.
Do not use the control solution beyond the expiry date or 3 months
after first opening. Write the opening date on the control solution
vial and discard the remaining solution after 3 months.
It is recommended that the control solution test be done at room
temperature 20°C to 25°C (68°F to 77°F). Make sure your control
solution, meter, and test strips are at this specified temperature
range before testing.
Shake the vial before use, discard the first drop of control solution,
and wipe off the dispenser tip to ensure a pure sample and an
accurate result.
Store the control solution tightly closed at temperatures between
2°C to 30°C (35.6°F to 86°F). Do NOT freeze.