Alternate Configuration:
THMS600-PS
Thank you for purchasing the THMS600-PS
Temperature Controlled Pressure System. Please
take the time to read through the manual as it will
help you to make the most out of the equipment.
The stage and its accessories are packed in the
protective hard plastic case. Its recommended
that you use this case to store the stage when it is
not in use for a long period of time
The THMS600-PS has specifically been designed
for use at 14bar. When in use, the stage must not
be taken above this pressure. There is a safety
pressure valve that releases gas pressure above
14bar. At 14bar of pressure the stage tempera-
ture range is: 100°C to 500°C
The THMS600-PS is designed for use with
pressurised air. For use with other gases,
please contact support@linkam.co.uk
THMS600-PS Stage Specifications
Maximum temp at atmospheric: 600 °C
Minimum temp at atmospheric: -196°C
Maximum pressure: 14 bar
Maximum Temperature at 14Bar: 500°C
Minimum Temperature at 14Bar
(with LNP95 System): -100°C
Maximum Heating Rate: 150°C/min
Maximum Cooling Rate: 30°C/min
Aperture hole size: 1.3mm
Stage Body Dimensions: 160x80x24mm
Weight: 1.2Kg
Objective Lens WD: 8.3mm
Condenser Lens WD: 14.5mm
THMS600-PS Pressure Systems
THMS600-PS Pressure System:
The system consists of a THMS600-PS stage, and
a T95-LinkPad System Controller.
Low Temperature THMS600-PS System:
The System has an additional LNP95 Liquid
Nitrogen Cooling Pump System and LINK Tem-
perature control software.
The stage is mounted onto the microscope by
using specific stage clamps or an adaptor plate.