Adjusting the Probe
Probe orientation:
The probe has a magnetic base (1) and can be
placed in any orientation inside the HFS600E-
PB4/PB2 stage.
The circular metal ring (2) holds the magnetic
foot of the Probe in place.
Adjust the length of the needle:
Pull end of the needle to the desired length and
use a wire cutter to cut the needle shorter.
Bend the tip of the needle:
Use a pair of flat pliers to bend the tip of the
needle.
Note: this action can only be done once. Tung-
sten metal is brittle and bending it again will
break the needle.