Installation Manual WTS3 Series II
M013610b.doc 4
SET UP : SWITCH SW3 SELECTION
S
ELECTION OF THE CURRENT RANGE
Select switch SW3-5 according to the following table to select the output current range 4 - 20 mA
or 0 - 20 mA :
SW3-5 Output signal range
S
ELECTION OF ANALOG FILTER
The device provides for the possibility of choosing between two types of damping of mechanical
vibrations or electrical noise. Select the switch SW3-4 for the desired choice of the filtering of the
analog signal as shown in the following table :
S
ELECTION OF COARSE AMPLIFICATION
It is possible choose from three fields of signal amplification of the load cells. Select the switch
SW3-1, 2, 3 for the choice of the desired range as indicated in the following table :
(Coarse amplification)
From 0,4 to 3,1 mV/V of useful signal
From 0,2 to 1,5 mV/V of useful signal
Special version – Customer application
ATTENTION :
• Our technicians are available for any requests for customization of the card with
fields other than standard.
• SW3-6 : not used.