Instructions for use Lojer Modux nursing bed / Version 2.5
Copyright © Lojer Oy, 2022 20
6.6 Sleeping surface raising and tilting
The sleeping surface’s lifting mechanism is powered by an electric motor. The
height of the nursing bed can be adjusted between 30–69 cm and with
optional dual locking wheel systems between 35-74 cm.
When adjusting the bed, always make sure that there are no people or
objects under the bed or between the structures (risk of squeezing).
The nursing bed should always be left in its lowest position to minimize
the risk of falling or crushing and increase user safety.
Raise or lower the sleeping surface by pushing the appropriate button on the
hand control unit.
1 Raising the bed surface
2 Lowering the bed surface
Image 27. Raising and lowering the sleeping surface
Tilt the sleeping surface to Anti-Trendelenburg -position (head end of the bed
raised) by pushing the appropriate button on the hand control unit (Item 1 in
Image 28).
Reset all the sections of the sleeping surface (back, thigh and leg sections,
sleeping surface and tilt) by pushing the appropriate button on the hand
control unit (Item 2 in Image 28).
1 Tilting the sleeping surface to Anti-Trendelenburg
2 Reset all the sections of the sleeping surface
Image 28. Adjustments for the Anti-Trendelenburg and reset -function