Instructions for use Lojer Modux nursing bed / Version 2.5
Copyright © Lojer Oy, 2022 37
Service
Lojer beds need little maintenance, because the bed’s electric devices are
maintenance free and its moving parts are lubricated. When the instructions
for normal use and cleaning are followed, the lubrication should not wear off.
However due wearing caused by use of the bed, maintenance lubricating
should be conducted as instructed annually.
The expected lifetime of the bed and its’ accessories is at least 10 years,
provided that the bed and accessories has been used and maintained according
to the manufacturer’s instructions. Faulty bed should not be used. All original
spare parts for the bed are available from the manufacturer, and we recommend
replacing any broken or worn parts.
Before starting servicing, make sure that the mains lead is
disconnected and that the functions have been locked. Check by using
the hand controller.
Service technicians must always read the instructions for use before
carrying out service/repair work, and they must obey safety instructions
when working.
The Holder is responsible to ensure that only trained or adequate
qualified person is authorized to service or repair the device.
Maintenance work carried out by an unauthorized person may cause
personal injury or damage to the device, for which the manufacturer is
not responsible.
Allow the electrical components below the bed mattress to cool down
before servicing or cleaning them. Maximum contact time of
components consecutive on bare skin < 1min.
Use only original spare parts, available from the manufacturer, and
follow the manufacturer’s guidelines.
Check that the bed is working properly after each maintenance/repair
procedure.
Electrical and mechanical components (excluding the accessories
referred to in chapter the operating instructions 'Installation and use of
accessories') may only be installed by service personnel authorized by
Lojer service (vocational upper secondary qualification in machine and
metal work or equivalent and valid SFS 6002 training in occupational
safety in electric work) or personnel otherwise qualified for service of
medical devices.
Document all maintenance work. In accordance with our warranty
terms and conditions, insufficient maintenance may also cause the
warranty to lapse.