Electromagnetic compatibility (EMC)
12.1 General information on EMC
Medical electronic devices must be installed and used in accordance with the
electromagnetic compatibility (EMC) information described in this manual.
Portable radio frequency communication devices may affect the operation of
this device.
Other devices may be disturbed by EMC radiation that is even slightly above
the reference value indicated in the standard. To determine whether the
resulting interference is caused by this device, start and stop this device. If, as
a result, interference in other devices disappears, this device is the cause of
the detected interference. In such rare cases, interference may be reduced or
eliminated by the following means:
Transfer this device and other devices to a different location, move
them to a different position or a different distance from each other.
Ensure that the device is not exposed to electromagnetic radiation
exceeding applicable norms. Portable and mobile RF communications
equipment can affect to the device.
Use of accessories, transducers, and cables other than those specified
or provided by the manufacturer of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation.
Press only one button at a time from hand control unit. Pressing
multiple buttons simultaneously can lead to improper operation of the
equipment in the surroundings.
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such
use is necessary, this equipment and the other equipment should be
observed to verify that they are operating normally.