Failure of internal units
261
Measures for failure of the expiratory flow measurement
The device continues to ventilate in the current
ventilation mode (no display for MV and V
Te
, only
inspiratory flow and volume curve).
Check the expiratory flow sensor for soiling and
damage at the next available opportunity. If
necessary, change the expiratory flow sensor.
Execute a system test at the next available
opportunity.
If you are unable to fix the error yourself, make a
note of the error number and notify a
Löwenstein Medical-authorised technician.
Pressure measurement failure
Reaction of the system to failure of the pressure measurement
Possible messages:
Sensor fail, only MAN/SPONT possible
The system switches automatically to the
MAN/SPONT form of ventilation.
The buttons to select the forms of ventilation are
inactive.
Measures for pressure measurement failure
The patient can continue to be ventilated with the
ventilation bag.
If you are unable to fix the error yourself, make a
note of the error number and notify a
Löwenstein Medical-authorised technician.
Pressure measurement failure!
Too-high ventilation pressures lead to lung damage
The patient can continue to be ventilated with the ventilation bag.
Set an alternative ventilation pressure measurement.