7. H
maintenance
Cleaning
the main head
Cleaning
the main head
7. H
maintenance
3736
Cleaning the Keymodule:
1. Turn the Keymodul and clean the rollers
on either side. Manually turn the rollers
so as to gain access to the entire surface
area.
2. Turn the Keymodule again and refit
the sealing valves as shown on the photos.
3. Do not forget to clean the sealing joint
between the Keymodule and head body.
4. Check that the electrical contacts are
clean and dry and proceed in the reverse
sequence to refit the Keymodule onto the
main head.
Now check that the head is operating
correctly. To do this, operate the head in the
palm of your hands to ensure the correct
sealing of the treatment chamber and
visually inspect the movement of the valves
and that of the rollers.
WARNING: if the Keymodule is not fitted to
the main head, if it is incorrectly positioned
or even faulty, the main head screen will
display the icon alongside:
Check the assembly of the Keymodule.
electrical contacts
sealing joint between
the head body and the
keymodule
For hygiene reasons, the treatment heads
must be cleaned after each use, using
antiseptic wipes with a bactericidal and
fungicidal solution.
Special attention must be paid to those
parts coming into contact with the patient.
Cleaning the tightness valves:
1. Release the Keymodule
as shown
on the photos.
2. Make sure that the cursor is in the top
position.
3. Bring the roller to the center.
4. Remove the corresponding sealing valve by
handling it as shown on the photo.
Repeat this operation for the other valve.
5. Then clean the valves and their location.
Replace the valves after:
100 hours’ operation.
!
Do not use aggressive products, such as acetone, trichloroethylene or alcohol at 90°, or abrasive sponges.
!
Do not use aggressive products, such as acetone, trichloroethylene or alcohol at 90°, or abrasive sponges.