Channel Setup
MXD70 Electrodeless Conductivity
Setup and Operating Guide
- 17 -
Channel Setup
TDS Factor
When TDS is selected as the operating units the
instrument will display the conductivity as “ppm”
using a factor which can be adjusted between 0.50
and 0.90.
/ – Increase / Decrease Digit
– Select Next Digit
EXIT
– Cancel
– Save Value
Temperature Input Sensor
Select the input channel’s temperature sensor type
for use with the sensor measurement’s automatic
temperature compensation.
If a temperature sensor is not connected to the input
channel then this menu item should be set to
disabled, else temperature input error messages will
be shown.
Note. Even when disabled is set a manual
temperature compensation can be used.
/ – Select Option
EXIT
– Cancel
– Save Selection
Temperature Units
Sets the temperature units used.
/ – Select Option
EXIT
– Cancel
– Save Selection