Registers
- 68 -
Modbus RS485 Interface
Operating Guide
Registers
Setpoint Configuration Continued
8046 Get/Set Setpoint 1 Cleaning
Recovery –
Seconds
INT Seconds element for
Cleaning Recovery Time
*(Only available when
Cleaning Mode set to
Offline)
0 to 59 minutes
8126 Setpoint 2
8206 Setpoint 3
8286 Setpoint 4
8366 Setpoint 5
8456 Setpoint 6
8047 Get/Set Setpoint 1 Cleaning Delay INT Cleaning Delay 1076 = Yes
8127 Setpoint 2 1077 = No
8207 Setpoint 3
8287 Setpoint 4
8367 Setpoint 5
8457 Setpoint 6
8050 Get/Set Setpoint 1 Unit Alarm
Condition*
INT Alarm Mode
*(Only Available when
Channel is set to Unit
Alarm)
1137 = Disabled
8130 Setpoint 2 1138 = Sensor Error
8210 Setpoint 3 1139 = Dose Alarm
8290 Setpoint 4 1140 = Calibration
8370 Setpoint 5 1141 = Offline
8460 Setpoint 6 1142 = Any Error
1143 = Cleaning
1144 = Calibration Due
1145 = Gain error
1146 = Power Failure