Loading...
9. Cleaning
Cleanthebathliftasfrequentlyasyoucleanyourtub,usingnon-abrasivecleaners.*Ifused
communally,werecommenddisinfectingthebathliftaftereachuse.SOLVENT-basedcleaners,such
as white spirit, should NOT be used.
i) Todisinfectthebathlift,useasoftbrushorclothandwipedownwithasterilizingsolutionor
anti-bacterialcleaner.Rinsetheproductwithcleanwater,thenwipeanddryusingacleancloth.
ii) Bath/medicinaloilsandothersolutionsareeasilyremovedwithwashing-upliquid.Donotallow
oilsorotheradditivestobuildup.Theyshouldberemovedfrequentlyotherwise,overthecourse
oftime,theymayleadtodegradationofthesuctionfeetand/ordiscolorationofthebathlift.
iii) Thebathliftcanbecleanedinanautomatedwashingmachine(toamaximumtemperatureof
73°C/163°Ffor3minutes),howeverthebatteryMUSTberemovedandwashedbyhand.The
batterymustnotbesubjectedtohightemperaturecleaning.
iv) Donotcleanorrinsetheactuatorunderaconstantowofwater,e.g.runningfaucet,ortotally
submergeitinwater.
v) Ifyourbathliftisttedwithsoftcovers,ensureyouremovethemonaregularbasistocleanthe
bath lift surfaces.
Itisalsoadvisabletocleanthetubsurfaceunderthesuctionfeetfromtimetotime.
*Cleaningagentsshouldbeusedinaccordancewiththeirmanufacturersguidelines.
NOTE:
Iftheactuatorisdetachedfromthebathliftforcleaning,ensurethebatteryisalways
disconnectedfromthehandcontrol.
10. Maintenance and servicing
RepairsandservicingmustbeentrustedtoaLumex®Splashauthorizeddealerorengineer.
ThebathliftshouldbeinspectedannuallybyaLumex®Splashauthorizeddealerorengineer,who
willcheckthattherearenoloosettingsoranyotherevidenceofmechanicaldamage.
Intended life - bath lift
ThedurabilityofthebathlifthasbeentestedinaccordancewithBSEN10535:2006(Hoistsforthe
transferofdisabledpersons-requirementsandtestmethods). If serviced and used in accordance
withtheoperatinginstructions,thebathliftandallparts(excludingbattery)willachieve11,000
liftingcyclesor5yearsuse,whicheveristhesoonest.
Intended life - battery
It is essential that the operating instructions with regards to battery use and charging is adhered
to.Thebatterylifeexpectancyis500chargecyclesor1yearsuse,whicheveristhesoonest.We
recommendthebatterybereplacedaccordinglyinordertoobtainthebestperformancefromyour
bath lift.
11