2. SPECIFICATIONS
Display LCD size : 29 mm x 28 mm.
Dual function LCD
Circuit Custom one-chip of microprocessor LSI
circuit.
Measurement Dissolved Oxy
g
en 0 to 20.0 m
g
/L ( liter ).
& Ran
g
eAir Oxy
g
en 0 to 100.0 %.
* For reference only
Temperature 0 to 50 .
Dissolved Oxy
g
en 0.1 m
g
/L.
Resolution Oxy
g
en in Air 0.1 % O2 .
Temperature 0.1 .
Accuracy Dissolved Oxy
g
en ± 0.4 m
/L.
(23± 5 ) Oxy
g
en in Air ± 0.7% O2.
* For reference only
Temperature ± 0.8 /1.5 .℃℉
Sensor The polaro
g
raphic type oxy
g
en probe with
Structure an incorporated temperature sensor.
Temp. 0 to 50 ,
Compensation
* Automatic adjustment
Memory Records Maximum, Minimum readin
g
s with
Recall recall.
Power of
f
Auto power off saves battery life, or manual
off by push button.
Data Hold Freeze the display reading.
Sampling Time Approx. 1 second.
of display
Operating 0 to 50 .
Temperature
Operating Less than 80% RH.
Humidity
Power Supply DC 1.5V battery ( UM-4/AAA ) x 4 PCs.
Power Current Approx. DC 6.2 mA.
Weight 176 g/0.39 LB.
* Include probe and batterry.