D. Dissolved oxygen
Oxygen Optional,
Probe The polarographic type oxygen probe with
Measurement Dissolved Oxygen 0 to 20.0 mg/L ( liter ).
& Range Oxygen in Air 0 to 100.0 %.
Temperature 0 to 50 .
Resolution Dissolved Oxygen 0.1 mg/L.
Oxygen in Air 0.1 % O2 .
Temperature 0.1 .
Accuracy Dissolved Oxygen ± 0.4 mg/L.
(23± 5 ) Oxygen in Air ± 0.7% O2.
Temperature ± 0.8 /1.5 .℃℉
Probe Temperature 0 to 50 ,
Compensation Automatic
& Adj. Salt 0 to 50 % Salt
Height ( M. T.) 0 to 8900 meter
Probe Weight 335 g/0.74 LB ( batteries & probe included )
Probe Size 190 mm x 28 mm Dia. ( 7.5" x 1.1" Dia. )
Optional * Oxy
g
en probe....................................
.
OXPB-11
Accessories * Spare Probe head with Diaphragm set
........................................................
.
OXHD-04
* Probe-filling Electrolyte.................... OXEL-03
@ Above specification tests under the environment RF Field Strength
less than 3 V/M & frequency less than 30 MHz only.