User Interface Commands
4-24 FOT Lab Kit User’s Manual
UN Unit of Measure
The UN command is used to report or select the serial-interface data
output units. Three units are currently supported: Kelvin, Celsius
and Fahrenheit. The analog outputs are mapped directly from the
serial output units.
NOTE For legacy reasons it is important to select the output units first
before adjusting any other system parameters. Units should not be
changed after other parameters are set or after calibration has been
performed. If you change units the device will need to be recalibrated.
Information Request:
Syntax: <Esc>UN ? <CRLF>
Response: UN = units
Reports the current units/data type being transmitted through
the user interface.
CELSIUS Temperature in degrees Celsius
KELVIN Temperature in degrees Kelvin
FAHRENHEIT Temperature in degrees Fahrenheit
Parameter Modification:
Syntax: <Esc>UN = units
Range: CELSIUS
KELVIN
FAHRENHEIT
Examples:<Esc>UN = KELVIN
<Esc>UN = K