User Interface Commands
4-26 FOT Lab Kit User’s Manual
Standby Mode
In Standby mode sampling and temperature reporting are
disabled and the device simply waits for additional commands
from the PC host.
Calibration Mode
In Calibration mode sampling and temperature calculation are
still active but the temperatures, rather than being reported are
used to calculate new calibration factors.
Of the one-letter codes described below some are valid only in
certain operational modes or have different functionality in different
modes. For this reason the sequence in which these commands are
entered is significant. Please read the command descriptions
carefully to obtain optimal results and avoid confusion.
Example of a typical Remote Control mode session:
■ Place device into Standby mode using CTRL+T.
■ Place device into Remote Control mode using CTRL+E.
■ Send the Run command CTRL+R.
■ Send the Initiate Measurement command CTRL+I. Wait for
CTRL+I to be echoed.
■ Send the Report Measurement command CTRL+Q. The
report will be transmitted.
■ Repeat steps 4 and 5 as desired.
■ Return to from Remote Control mode to Standard mode using
CTRL+D.
■ Re-enable periodic temperature reporting using CTRL+R.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com