User Interface Commands
4-28 FOT Lab Kit User’s Manual
CTRL+F Flush the Measurement Buffer
Discard any data currently stored in the measurement buffer. Use
this command when you want to discard old data to prevent it from
affecting new data. For example you may have just installed a new
probe and want to discard any garbage data still in the input buffer.
Old data will, of course cycle out on its own but may take several
seconds depending upon the combination of system parameters
you have selected. This command simply speeds up that process.
The CTRL+F character will be echoed back to the host.
CTRL+I Create a Data Report
This command is valid ONLY when the device is in Remote Control
mode (CTRL+E command).
The CTRL+I command initiates data sampling and the creation of
one temperature report. When the report is ready the device will
echo the CTRL+I character. You may then request the report be
transmitted using the CTRL+Q command.
CTRL+K Initiate Auto-Calibration Sequence
This command is valid ONLY when the device is in Standard mode.
The CTRL+K command invokes the auto-calibration sequence for
all active inputs. When the sequence is complete the device will
report the status of the operation as shown with the status indicator
immediately following the colon (':') and the calibration
temperature. Please see Appendix A, “Calibration Procedure” for
further details.
1:C 348.15 K 2:C 348.15 K 3:C 348.15 K 4:C 348.15 K