User Interface Commands
4-30 FOT Lab Kit User’s Manual
CTRL+T Enter Standby Mode
This command is valid ONLY when the device is in Standard mode
(CTRL+R command) or Remote Control mode (CTRL+E
command).
The CTRL+T command causes the device to enter Standby mode (if
not already in Standby). Standby mode can be thought of as a low-
power mode in that the lamps are turned off and temperature
reporting disabled. If command is successfully executed the
CTRL+T character will be echoed back to the host, else a question
mark ('?') will be returned.
CTRL+X Reset the Instrument
The CTRL+X command is used to perform a software reset of the
FOT Lab Kit device. This software reset is exactly equivalent to
cycling power to the unit.
All parameters will be loaded from NvRam exactly as they are on
power up.
If the parameters in any section of NvRam have been corrupted,
factory defaults will be loaded for that section and a message will be
displayed informing the user which section(s) have received the
factory default values.