Temperature Data Reporting
5-2 FOT Lab Kit User’s Manual
The following is an exact breakdown in both ASCII and hexidecimal
of a report where input channels 1 and 2 are active and DF = ABBR.
Space characters are replaced by the underscore character '_' for
visual clarity.
1. The first two characters on the line are spaces.
2. The next character is the input channel number followed by a
colon and one space.
3. The temperature value follows which is four places to the left
of the decimal, the decimal point and two places to the right of
the decimal. The value is right-justified with leading spaces as
necessary.
4. The temperature is followed by one space and the units
designator.
5. Items 1 - 4 are repeated for succeeding channels.
6. If the temperature units are Celsius or Fahrenheit the
temperature may be below 0.00, in which case the minus
character ('-') is displayed as the first digit of the channel's
temperature value (e.g.__1:_-_41.50_C).
7.
The line is terminated by the carriage-return / linefeed pair.
Example showing a report line with all four channels active.
1: 224.39 C2: 224.51 C3: 224.22 C4: 224.19 C
Example showing a report line with channels 1 and 2 active, then 1,
3 and 4 active with DF = FULL.
1: 224.39 C2: 224.51 C3: 4:
1: 224.39 C2: 3: 224.22 C4: 224.19 C
Example showing a report line with channels 1 and 2 active and DF
= ABBR.
1: 224.39 C 2: 224.51 C
_
_ 1 : _ _ 2 2 4 . 3 9 _ C _ _ 2 : _ _ 2 2 4 . 5 1 _ C
20 20 31 3A 20 20 32 32 34 2E 33 39 20 43 20 20 32 3A 20 20 32 32 34 2E 35 31 20 43 0D 0A