A-6 FOT Lab Kit User’s Manual
10.
Start the automatic calibration sequence using the CTRL+K
command. You may abort the procedure anytime by pressing
CTRL+A command.
11. Wait for a few seconds for the calibration to complete, then
review the output data for errors (see above).
12. After the unit has finished the calibration of channels 1 and 2,
then calibrate channels 3 and 4 with the following sequence.
13. Disable channels 1 and 2 and enable 3 and 4 with the following
command <ESC> PS=3,4 ENTER.
14. After you have enabled channels 3 and 4, disable the
calibration again with the command <ESC> CC=D ENTER.
15. Place the probes attached to channels 3 and 4 in the 100
o
C
temperature bath and wait until the temperature is stable (5 to
15 minutes).
16. Start the automatic calibration sequence using the CTRL+K
command. You may abort the procedure anytime by pressing
CTRL+A command.
17. Wait for a few seconds for the calibration to complete, then
review the output data for errors (see above).
18. After the calibration of channels 3 and 4 is complete, enable all
four channels at once with the command <ESC> PS=1,2,3,4
ENTER.
19. If all four channels are reading correctly, then save the setting
with the following command <ESC> SV ENTER.
If there are no errors the new correction factors have been
calculated and will take effect immediately.
NOTES Do not issue any commands to the device for at least 10 seconds
after the calibration routine has finished. This allows the device time to
reinitialize the data-sampling sequence.
If an error is encountered during the procedure and error message
will be displayed for the failed channel(s). In this case revisit all
steps of the setup and calibration to determine the cause of the
error and run the procedure again.