8.5.2 Synchronized oil-UV feeding
Manual Oil-UV ■
Synchro. Oil-Uv □
Use UP-DOWN buttons to select Synchronized Oil-UV. Confirm by pressing ENTER.
ISO oil:
46 ■ 100 □ 150 □
Use UP-DOWN buttons to select oil ISO parameter. Confirm by pressing ENTER.
Specify oil qty
g 10
Use UP-DOWN buttons to select oil quantity to be introduced. Confirm by pressing ENTER.
Synchronized Yes
UV feeding No
Select using UP-DOWN buttons. Confirm by pressing ENTER.
If NO option is selected, the station will suggest going into the next phase.
If YES option is selected, the following message is displayed:
Contrast qty
to be introd. g 10
Use UP-DOWN buttons to select contrast quantity to be introduced. Confirm by pressing
ENTER. The station will suggest going into the next phase.
Note: make sure oil and contrast in the tanks is sufficient to perform the feeding
operation.
Synchronized feeding is done with the accuracy of +/- 15 g (at 20 to 30°).
If synchronized injection option is selected, changing the oil quantity is still possible during
vacuum generation phase.
press and hold ENTER button for 3 seconds – planned quantity of oil to be fed is displayed.
Use UP-DOWN buttons to select oil quantity. Confirm by pressing ENTER.
Display will return to showing the remaining vacuum time.