EasyManuals Logo

Maico MA 25 Operating Instructions

Maico MA 25
25 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #10 background imageLoading...
Page #10 background image
Operating Instructions MA 25 / MA 25e
operating instructions_8012238-4_MA25_E_13a 8 8012238-4 10/13
3.2 Special functions of MA 25e
Talk Forward: Talk Forward is activated by holding down the
Hearing Level db dialer.
While holding down the Tone Switch the talk forward level can be adjusted.
Hughson-Westlake: The following MA 25e F-key functionalities can be
accessed by pressing the frequency dialer.
F1 - Delete all thresholds stored in the internal memory of the MA 25e.
F2 - Store a Not Heard threshold point.
F3 - Display the L/R thresholds stored in the internal memory of the MA 25e.
F4 - Start the Hughson Westlake (HW) automatic test procedure. Please refer
to the chapter 5 for instructions about how to setup the HW test.

Other manuals for Maico MA 25

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maico MA 25 and is the answer not in the manual?

Maico MA 25 Specifications

General IconGeneral
BrandMaico
ModelMA 25
CategoryMedical Equipment
LanguageEnglish

Related product manuals