18/05/2022
2.5 Establishing a Connection to a PC, Printer or an External Audio Device
Infection of the device or the software used with the device can
lead to system failure and data misuse.
Ensure that your PC is adequately protected against cyber-attacks.
To transfer data to a PC, establishing a PC connection via USB is required. If the MA 42
is used with office equipment that is not a medical device itself (see Table 8PC
Connection 1), make sure to establish the PC connection in one of the following ways
(see, PC Connection 2, 3 or 4).
In case an external audio device shall be connected proceed accordingly to PC
Connection 3 (if used with an external audio device with power supply) or 4 (if used with
a battery driven external audio device).
Make sure you use only office equipment with the device that
is a medical device itself or meets the requirements of
IEC 60950. If a non-medical device is used within the patient
environment (1.5 m from patient as defined in IEC 60601)
a voltage transformer must be used (exception: a battery driven
laptop is used).
Table 8 PC-Connections
PC CONNECTIONS
PC Connection 1:
Medical device Medical device
PC Connection 2:
Medical device Non-Medical device
PC Connection 3:
Medical device Non-Medical device
PC Connection 4:
Medical device Laptop (battery-driven)
WARNING
WARNING